Applications
Non patient contact medical
General electronics
IT equipment
RPT-75 Series

Case No: 8015FM
Dimensions: 127*76.2*31 mm (L*W*H)

Features
Universal AC input/ Full range
Low leakage current <100μA
Protections: Short circuit, Overload and Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
ANSI/AAMI EDS60601-1/IEC60601-1/EN60601-1 Medical
safety approval
Fixed switching frequency at 65KHz

Part Number

Voltage (max)

Current (amps)

Power (watts)

RPT-75A

5V

6A

68.5W

12V

3A

-5V

0.5A

RPT-75B

5V

6A

72W

12V

3A

-12V

0.5A

RPT-75C

5V

6A

72W

15V

2.3A

-15V

0.5A

RPT-75D

5V

5A

73W

24V

2A

12V

1A

RPT-7503

3.3V

6A

62W

5V

6A

12V

1A

Description
Complying with both medical and industrial agency standards, these units can deliver their full rated output of 75W with only free air convection and up to 100W by providing 23.5CFM forced air. Featuring only 200µA of leakage current at 264VAC, they can be equipped into non-patient contact medical equipment and still comply with Class B EMI regulations. Standard functions include short circuit, over load, over voltage protections, and the operating temperature can be as high as 70°C under forced air convection. For a greener world, RPS/D/T-75 are all RoHS compliant and the no load power consumption for single output models is as low as 0.75W so that they can be used in powering electronic equipments requiring green mode.